TM-2915

NIBP Module

A&D has developed the non-invasive blood pressure (NIBP) module TM-2915 with a new algorithm.
TM-2915 supports a wide range of devices from medical monitors to therapeutic devices.
Its flexible design enables use with various host devices.

Accessories & Options

■ Standard Accessories
 Adult cuff
 Air hose (2m)
 Cuff connectors
 Silicon tube (300mm)

■ Options
 Child cuff
 Small cuff
 Large cuff
 Air hose (1.5m)
 Disposable cuff clothes (2 sets)

*The cuff connector and air hose length for the host device can be customized as necessary.

Product Information

Features
Advanced measurement algorithm
Compact, quiet, low-heat design
Includes the well-established ECEV* ( Electric Control Exhaust Valve)
Covers measurement from newborns to adults
Specifications
Measurement MethodOscillometric Method
Pressure DetectionFor MeasurementPressure sensor for capacitance method
For Safety CircuitPressure sensor for capacitance method
Measurement RangePressure0-300 mmHg
Pulse40-200 bpm (adult)
40-240 bpm (newborn)
Measurement PrecisionPressure±3 mmHg
Pulse±5 %
Pressure MethodPumpRolling Pump
Pressure Performance10s or less / 0→300 mmHg
(At shipping)(Power voltage 5.00V, 500cc capacity connection)
Exhaust MethodConstant ExhaustBy solenoid valve control (ECEVII)
Quick exhaust by opening solenoid valve for constant exhaust or quick exhaust valve.
Power SourceFor Circuit (Vcc)5±5 % [v]/0.1[A]
For Mechanical Power (Vp)5±5 % [v]/1.0[A]
Serial CommunicationVoltage LevelTTL (3.3V) *RxD up to 5V possible
Trans, MethodStop-start synchronization, duplex transmission
Trans. SpeedAt shipping from factory, selectable from 4800 / 9600 / 19200 bps
Data Configuration1 start bit, 8 data bits, no parity, 1 stop bit
Operation EnvironmentTemperature+10 to +50°C
Humidity85%RH or less (with no condensation)
Pressure700 to 1060hPa
Storage EnvironmentTemperature-20 to +60°C
Humidity95%RH or less (with no condensation)
Pressure500 to 1060hPa
External Dimensions60(H) x 39(W) x 104.5(D)
Weight280±5% [g]
Brochure
TM-2915