Multifunctional Test Instrument
All safety relevant, characteristic electrical values can be measured and documented for portable and stationary medical devices with the menu-driven SECULIFE | ST. And you even have the choice between fully automated or manual test sequences.
The SECULIFE | ST detects mains supply power, protection class and mains connection errors at the device under test. The test sequence is significantly simplified by the automatic mode.
• Select a regulation.
• Press the start key.
• Read the results.
Complete function tests can be also conducted with mains power on large, high-power devices with the integrated test socket.
The SECULIFE | ST is equipped with an operator safety switch.
The SECULIFE | ST is the only measuring instrument available on the market which has been designed to test medical devices operated with 3-phase electrical current. AT3-IIIE and AT3-IIS 3-phase current adapters can be furnished to this end as optional accessories.
Possible areas of use and measurements performed with the SECULIFE | ST are expanded with numerous multimeter functions. The performance of individually configured single measurements is thus made possible.
As a standard feature, the SECULIFE | ST is equipped with an RS 232 port for transferring measurement and test reports to a PC or a printer.
The instrument can be optionally expanded to include a printer-memory-interface with GOSSEN METRAWATT’s external (P)Si module. The SECULIFE | ST is compatible with PS 3, visualFM, Fundamed, MD Data, PC doc Word/Excel, PC doc Access and other software.
Product Information
| Features |
|---|
| Protective conductor resistance with 200 mA |
| and 10 or 25 A test current |
| Insulation resistance |
| Earth leakage current |
| Contact current (device leakage current) |
| Patient leakage current, AC and DC |
| Patient auxiliary current |
| Equivalent leakage current method |
| Direct measurement method |
| Differential current measurement |
| Up to 10 application parts (2 mm) can be connected and individually assigned to groups |
| Internal memory for 125 tests |
| RS 232 port |
| Mains connection: 110 V / 60 Hz, 230 V / 50 Hz |
| Extensive accessories for testing 3-phase devices |
| Safety for the inspector in case of mains errors or increased residual current due to power-off |
| HV test (option) |