Tecotherm Neo

Total Body Cooling and Warming

Tecotherm Neo utilizes an innovative servo controlled design with instant feedback, monitoring the infant’s temperature every 2 seconds and making minute changes to the cooling fluid to ensure that the infant’s temperature remains stable. Simple alarms on the Tecotherm Neo alert staff should the temperature deviate more than 0.5°C from the set temperature. The large colour display shows the temperature clearly and colour changes alert nursing staff to any problem.

With several new features, the Tecotherm Neo can be programmed for a complete treatment cycle. The user can set the target temperatures, duration and even the rate of re-warming / cooling, thereby making it easy and simple to set up and use. Changes can be made at any time and all changes, set and measured parameters are recorded on to a memory card, every minute, for later analysis

Product Information

Features
3 modes of operation for maximum flexibility in use
Clinician can set and adjust parameters at the start and throughout treatment
Data storage for allowing simple analysis of temperature profiles
USB output to CSV
Large colour screen is easy to see across the room
Small / Lightweight portable device
Internal battery restarts from same settings in event of power failure
Smooth, linear rewarming
Compatible with reusable or disposable accessories
Optional use of skin probes
Maintains normothermia after rewarming
Specifications
Options: Cooling &Warming
Patient weight max: <50 kg
Dimension: 375 mm x 190 mm / 215 mm x 310 mm
Weight without accessories: approx 7.2 kg
Central cooling module: Thermoelectrically based module
Hydraulic circulation system
System pressure max. 0,5 bar
Flow rate without / with mattress 500 ml / min (shorted) / up to 300 in use
Internal fluid reservoir capacity approx. 250 ml
Circulating fluid Sterile Water
Connectors / Couplings Quick Disconnect Couplings
Fill up / Refill Fill up set
Electrical Paarameters
Supply voltage / Mains 100-130V and 200-240V, 50-60 Hz
Power consumption max. 350 W
Patient safety / alarms
Lower temperature alarm limit +10°C
Upper temperature alarm limit +41°C
Set temperatures, lower limit +12°C
Set temperatures, upper limit +39°C
Ambient conditions
Operation / Treatment (Ambient Temperatures) + 5°C to + 27°C
Operation / Treatment 10% to 75%, not condensating Relative Humidity
System safety
Protection class Class 1, Risk Class II b, (Type BF)
Standards: DIN EN 60601-1
DIN EN 60601-1-2
DI 60601-1-6
DIN EN 60601-1-8
DIN EN 60601-1-10
DIN EN 60601-2-35 E/F
Brochure
Techoterm Neo Brochure